(3) Connect the cap connector tubes to the main
hose using the male/female connectors.
(4) After all the tasks are complete, press Next.
f. Setting up fill cycle, for the setup wizard displays
(1) Check the hose and tube connections, then
Table 45.1 Recommended Initial Cap
Table 45.2 Determination of Cap Size
Cap Size Infant Weight Head Circumference
Chapter 45 ■ Brain and Whole Body Cooling 329
Fig. 45.5. Water cap (A), Water cap (B), and Water cap retainer (C) placement on the infant.
A: Rocking, massaging, and inverting the water cap to remove any air bubbles from the crevices.
(2) After the system has filled, gently rock, massage,
and invert the water cap to remove any air
bubbles from the cap crevices (Fig. 45.5 A).
(3) After the tasks are complete, touch Next.
g. Placing Caps and Shield, for the setup wizard displays
(1) Position the water cap (Fig. 45.5B) on the
infant’s head, with the opening anterior, centered on the hairline and with the forehead
(2) Position the ear cutouts above the ears.
(3) Place the ear straps under the ears.
(4) Route the scalp temperature sensor wire out of
the cap, over the infant’s forehead.
(5) Place the cap connector tubes in the cap tube
(6) Place the blue water cap retainer (Fig. 45.5C)
(7) Slide the chin strap of the water cap through
the ear straps of both water cap and the water
cap retainer, and fasten the right and left
(8) Place the large Velcro fastener on top of the
infant’s head, in front of the red cap tube, and
the small Velcro fastener between the cap
(9) Place the silver insulating cap (Fig. 45.6A)
(10) Secure the three Velcro fasteners on the inside
of the water cap retainer to the patches on the
(11) Secure the Velcro fastener of the insulating
cap in front of the cap connect tubes.
(12) Place the heat shield over the infant’s head,
aligning with the chin, and resting evenly and
directly on the mattress (Fig. 45.6B).
(13) After each task is complete, touch Cool to start the
system’s cooling timer and main display screen.
h. Once the rectal temperature reaches 35.5°C, switch
i. Set the servo control temperature on the radiant
j. Adjust the water cap temperature to maintain the
rectal temperature between 34°C and 35°C.
k. Continue cooling for 72 hours.
l. Commence rewarming after 72 hours and follow
instructions on the screen for automated rewarming
at 0.5°C/h. Switch off the cooling unit.